Place of Origin: | China |
Brand Name: | DahoMeter |
Certification: | FCC, ISO |
Model Number: | DH-DSC-500Q |
Minimum Order Quantity: | 1 set |
---|---|
Price: | Negotiable |
Packaging Details: | standard exporting wooden package |
Delivery Time: | 5-7 working days |
Payment Terms: | L/C, D/A, D/P, T/T, Western Union, MoneyGram |
Supply Ability: | 300 sets per month |
Temperature Range: | Room Temperature~500℃ | DSC Range: | 0~±600mW |
---|---|---|---|
Display Mode: | 15-inch LCD Touch Screen Display | Temperature Control Method: | Heating, Constant Temperature (automatic Program Control) |
DSC Differential Scanning Calorimeter for Plastic Rubber OIT Differential Scanning Thermal Analyzer DH-DSC-500Q
Introduction:
Differential scanning calorimeter is a touch screen type, which can perform glass transition temperature test, phase transition test, melting and enthalpy value test, product stability, and oxidation induction period test.
Standards:
GB/T 19466.2-2004/ISO 11357-2:1999 Part 2: Determination of glass transition temperature
GB/T 19466.3-2004/ISO 11357-3:1999 Part 3: Determination of melting and crystallization temperature and enthalpy
GB/T 19466.6-2009/ISO 11357-3:1999 Part 6: Determination of Oxidation Induction Time and Oxidation Induction Temperature
Technical parameters:
Temperature range | room temperature~500℃ |
Temperature resolution | 0.01℃ |
Temperature fluctuation | ±0.1℃ |
Temperature repeatability | ±0.1℃ |
Heating rate | 0.1~100℃/min |
Constant temperature time | recommended <24h |
Temperature control method | heating, constant temperature (automatic program control) |
DSC range | 0~±600mW |
DSC resolution | 0.01mW |
DSC sensitivity | 0.01mW |
Working power supply | AC220V/50Hz or customized |
Atmosphere control gas | nitrogen, oxygen (automatic switching of the instrument) |
Gas flow | 0-300mL/min |
Gas pressure | 0.2MPa |
Display mode | 15-inch LCD touch screen display |
Data interface | standard USB interface |
Parameter standard | Equipped with standard substances (indium, tin, zinc), users can calibrate the temperature by themselves |
Features:
1. The industrial-grade 15-inch touch screen is completely free from the constraints of the computer.
2. Electronic flowmeter, record flow more accurately.
3. Independent of the displayed touch interface, more flexible.
4. Unique furnace body sealing can provide good experimental conditions.
5. Automatically switch the two-way atmosphere flow, the switching speed is fast, and the stabilization time is short. At the same time, a protective gas input is added.
6. The software is simple and easy to operate.